Micro Series DC Voltage
&
Current Panel Meters
High accuracy, high read
rate, control outputs
|
Features
- Available in
three digit heights: 0.56" for panel mounting,
1.0" (Mighty-1) or
5.0" (Mighty-5)
for long-distance viewing.
-
0.2, 2, 20, 200 and 660 V ranges.
-
2, 20, 200 mA and 5 A ranges.
-
All ranges factory calibrated.
-
99.99% full scale accuracy.
-
4 1/2 digit resolution to ±19,999 for direct readout
in mV, V, mA or A.
-
1 Gohm
input impedance on 200 mV and 2 V scales.
-
Field scalable from front panel to ±99,999 for
use with external current shunts.
-
Up to 60 conversions per second.
-
Peak value display.
-
Selectable adaptive digital filter.
-
Selectable fixed zero or active least significant
digit.
-
5, 10 or 24 Vdc excitation supply.
-
External controls for reset, meter hold and decimal
points.
- Choice of isolated plug-in
options for control and computer interface:
dual
relays,
4-20 mA & 0-10
V analog output,
RS-232/485 serial I/O,
low voltage AC & DC power.
-
1/8 DIN
case environmentally sealed to NEMA-4X from front panel.
- 1/8 DIN
version certified to UL 3101-1, CAN/CSA-C22.2,
EN 61010-1 (CE Mark).
|

|
Description
Micro
Series DC voltage and current panel meters with a DC
signal conditioner board combine high accuracy with high
read rate and a wide range of isolated output options
for computer interface and control. Accuracy is 99.99%
of full scale ±1 count.
Used as a direct-reading
DC voltmeter, the Micro DC meter provides a full-scale
readout of ±20,000 counts and five full-scale voltage
ranges from 200.00 mV with 10 mV resolution to 660.0 V
with 100 mV resolution. The 200.00 mV and 2.000 V ranges
provide a high input impedance of 1 Gohm so as to
minimize the load on the voltage signal. The maximum
voltage which can be applied on the 20, 200 and 660 Vdc
ranges is 660 Vac.
Used as a DC ammeter, the
Micro DC meter provides a full-scale readout of ±20,000
counts and four full-scale direct-reading current ranges
from 2.0000 mA with 0.1 mA resolution to 5.000 A with 1
mA resolution. The 5.000 A range measures the IR drop
across a built-in 10 mW current shunt.
Use with
External Current Shunts

The Micro DC
meter can be used with external current shunts, which
typically produce 50 mV or 100 mV at their rated maximum
current. Scaling from millivolts to amperes for a
specific shunt
value is
easily accomplished from the front panel of the meter.
The scalable readout is five full digits up to ±99,999
counts. Since the voltage signal from a current shunt
can be noisy, the meter provides a selectable, adaptive
moving-average digital filter, as explained below.
All Ranges
Precalibrated
All voltage
and current ranges are calibrated at the factory, with
calibration factors for each range stored in an internal
EEPROM. This allows ranges to be changed without
recalibrating the meter.
High Read
Rate and Peak Capture
All Micro DC
voltmeters use Concurrent Slope (US Pat 5,262,780)
analog-to-digital conversion, which allows up to 60 or
50 conversions per second while integrating the signal
over a full power cycle. High read rate is ideal for
peak capture, real-time computer interface, and control.
The peak
value of the input signal is automatically captured and
may be displayed via a front panel pushbutton command or
a control signal at the rear connector since the last
meter reset. Other controls at the rear connector
include meter hold, meter reset, and decimal point
selection
Selectable
Signal Filtering
The
displayed readings and the data outputs can be
separately selected to be either unfiltered or filtered.
-
An unfiltered selection updates
after each conversion for fastest response.
-
A batch average filter selection
averages each 16 conversions for an update every 1/4
sec.
-
An adaptive moving average filter
selection provides a choice of 8 time constants from
80 ms to 9.6 s. When a significant change in signal
level occurs, the filter adapts by briefly switching
to the shortest time to follow the change, then
reverts back to its selected time constant. Another
choice is Auto, which provides an automatic time
constant selection based on the signal noise
characteristics.
Isolated Excitation Supply
5, 10 and 24
Vdc isolated excitation outputs are standard to power
external devices, such as transducers or transmitters.
In many cases, these outputs can eliminate the need for
an external supply.
Built-in Flexibility
Plug-in
isolated analog
output, dual
setpoint controller,
and
RS-232/485
serial communications boards can upgrade Micro and
Mighty Series
meters from a simple monitor to system interface and
control.
Micro Series
meters can be powered from
95-240 Vac ±10% and 95-370 Vdc, or
optionally from 12-34 Vac and 10-48 Vdc.
Any setup functions and front panel keys may be locked
out for simplified usage and security. The popular
1/8 DIN version with 0.56"
high digits is environmentally sealed to NEMA-4X (IP65)
from the front when panel mounted, and is
available with red or green LEDs.
|

|
Specifications
DC Voltage
|
DC Voltage
Range |
Resolution |
Input
Ohms |
Error at
25°C
±1 count |
200.00 mV
2.000 V
20.000 V
200.00 V
660.0 V |
10 µV
100 µV
1 mV
10 mV
100 mV |
1 G
1 G
1 M
1 M
1 M |
0.01% FS
0.01% FS
0.01% FS
0.01% FS
0.03% FS |
DC Current
|
DC Current
Range |
Resolution |
Input
Ohms |
Error at
25°C
±1 count |
2.0000 mA
20.000 mA
200.00 mA
5.000 A |
0.1 µA
1.0 µA
10 µA
1.0 mA |
100
10
1
0.01 |
0.01% FS
0.01% FS
0.01% FS
0.04% FS |
DC Voltage &
Current
|
Display |
Readout
Digit Height
Color
Range
Indicators |
5 digits, 7-segment
Available with 0.56", 1.0", 4.0" or 5.0" digits.
High brightness red. Green available
with 0.56" digits.
-99999 to +99999
-99990 to +99990 (selectable count by 10 with
rounding)
Minus sign,
2 red LED lamps |
|
A-to-D Conversion |
Technique
A-to-D Rate
Output Update
Display Update |
Concurrent Slope™
(Pat 5,262,780)
60/s at 60 Hz, 50/s at 50 Hz
56/s at 60 Hz, 47/s at 50 Hz
3.5/s at 60 Hz, 3/s at 50 Hz |
|
Accuracy |
Span
Tempco
Zero Tempco |
0.003% of reading/°C
0.1 counts/°C |
|
Noise Rejection |
CMR,
DC to 60 Hz
NMR at 50/60
Hz |
130
dB
90 dB with
min filtering |
|
Maximum Signal |
Max
applied
voltage
Current
protection |
660
Vac for 20, 200 and 600 V ranges, 125 Vac for
other
ranges
25x for 2 mA,
8x for 20 mA, 2.5x for 200 mA, 1x for 5 A |
|
Power |
Voltage, std.
Voltage, opt.
Frequency
Power isolation |
95-240 Vac ±10% and 95-370 Vdc
12-34 Vac and
10-48 Vdc
DC or 49-440 Hz
Safety-rated to 250 Vac, meter ground to earth
ground,
DC to 60 Hz, 4.2 kVp per High Voltage Test |
|
Excitation Output |
5
Vdc
10 Vdc
24 Vdc
Output isolation |
5
Vdc ±5%, 100 mA max
10 Vdc ±5%, 120 mA max
24 Vdc ±5%,
50 mA max
50 Vdc to
meter ground |
|
Connector Pin Assignments |
|
|
Environmental |
Operating Temp.
Storage Temp.
Relative Humidity
Protection |
0°C
to 60°C
-40°C to 85°C
95% at 40°C, non-condensing
NEMA-4X (IP-65) when panel mounted |
|